产品展示

Rosemount 848T Wireless Temperature Transmitter

作者:admin 发布日期:2011-4-18 13:24:31点击:1608

High Density Temperature Measurement 

Many important processes have temperature points within close proximity. Emerson delivers a proven and innovative transmitter family that provides a cost-effective, reliable way to access these measurements.


The Rosemount 848T High Density Measurement transmitters are designed specifically forFOUNDATION™ fieldbus and WirelessHART™ industry standards. These solutions provide a variety of benefits including:

Reduce wiring and conduit costs by measuring multiple points with a single transmitter

Minimize start-up time with fewer devices to install and commission

Avoid shutdowns with proactive maintenance by using device and sensor diagnostic

Rosemount 848T Wireless Temperature Transmitter 

A Wireless Solution for High Density Temperature Measurement Applications

Many users across the process industry need to monitor hundreds to thousands of temperature measurements in a plant to ensure efficient operations. The 848T Wireless monitors up to four independently configurable RTD, thermocouple, ohm, millivolt and 4 – 20 mA inputs allowing users access to more temperature measurements without any signal wire. Increased visibility to plant operations allows for proactive maintenance and monitoring of key assets such as motor bearings that drive pumps, fans and compressors, as well as heat exchangers, tanks, reactors and boilers.

Fully compatible with any WirelessHART™ network, the Rosemount 848T Wireless provides the same rich diagnostic data and reliable performance as other devices in the Smart Wireless family. It monitors many temperature measurements with one device ensuring accurate and stable measurements and continuously monitors measurement integrity. The device is self-powered and utilizes SmartPower™ to provide an intrinsically safe power module with long battery life.


Features & Benefits

Four independently configurable RTD, thermocouple, ohm, millivolt and 4 - 20 mA signals

Lowers installed cost by eliminating long wire runs to the control room

Dependable performance by enhancing measurement reliability for trustworthy data

-Mount close to process to keep sensor wiring short to minimize effect of electromagnetic interference(EMI)

-Transmitters compensate and filter noise and EMI for stable measurement performance

Simplified installation and configuration practices for faster start-ups

-Reduction of required loop drawings, eliminates cable trays and large wiring cabinets

-Enhanced user interface makes configuration intuitive for faster configuration and commissioning

Fully compatible with WirelessHART networks with greater than 99% data reliability

Eight user configurable alerts for advanced measurement monitoring

Temperature sensor diagnostics for continuous monitoring of sensor integrity

Type 4x, IP66 housings allows for installation in harsh process environments

Best ambient temperature compensation reduces measurement error in harsh applications, especially for  wide fluctuations of ambient temperature

Specifications

Reference Accuracy± 0.3°C (± 0.54°F) PT100 @ 20°C
± 0.70 °C (± 1.26 °F) for Type J Thermocouple
Stability±0.15% reading or 0.15 °C (0.27 °F), whichever is greater for 2 years for RTD
±0.15% of reading or 0.15°C (0.27 °F), whichever is greater for one year for thermocouples
Input2, 3, 4-wire RTD, thermocouples, mV, ohm, and 4-20 mA signals
OutputWirelessHART™
Power Module1 minute update rate – 6 years
10 minute update rate – 10 years
Update RateUser selectable 4 seconds to 60 minutes